Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Illustration 3-2:
The preferred installation method for the Innova LC Positioner or the Omega tables is through-
bolting. The through-bolting method can be used in all seismic zones. If through-bolting cannot be
used, use provided floor anchors instead.
Illustration 3-3: Through–Bolt Supplied (Slab Type Floor Construction)

3.3.2 Floor requirements when using provided floor anchors

The maximum pullout force per provided anchor was calculated assuming:
A concrete compression strength of 17.24 MPa at 28 days (which is the minimum required
compression strength).
Anchors installed to the required hole depth of 165.1 mm minimum, and
Chapter 3 Physical Requirements of Room
The Table baseplate or the System baseplate are mandatory to install the table
(patient support).
The Table must never be installed on grade.
Direction 5160944-13-1EN, Revision 2