Overview
While in Maintenance > RAID Storage, the RAID status can be found in the Raid pane.
Image 17–15
Storage state (reference 1)
RAID status (reference 2)

17.4.2 Initialize RAID storage

Location & access
Menu: Maintenance > RAID storage
Level : Administrator, Service technician
Inserting a 1TB hard disk in a 2TB system will result in a degraded condition. The 1TB drive will not
be adopted in the raid. Only 2 of the 3 drives are active and there is no raid protection in case one of
the drives fails. The system will work without redundancy.
Inserting a 2TB hard disk in a 1TB system triggers a rebuild and the 2TB drive is adopted in the raid
as a 1TB drive.
How to initialize
1.
While in Maintenance > RAID Storage, click Initialize in the Raid pane (reference 1)