Barco Freya User Manual page 189

Table of Contents

Advertisement

1
Image 9–3 White Point Calibration – White Point Tracking
The laser white point calibration will be reset. The default settings are programmed.
2.
Click Next (reference 2) to access the 2D measurement.
Image 9–4 White Point calibration - 2D measurement
3.
Measure the white coordinates (x, y) with spectrometer and enter in the x and y fields respectively
(reference 3).
4.
If 3D projection is required, click Next (reference 4) to access the 3D measurement. If not, press cancel to
exit.
2
Configuration - Color calibration
R5910032 /00
Freya
189

Advertisement

Table of Contents
loading

Table of Contents