Teledyne TIENet LaserFlow 360 Installation And Operation Manual page 83

Velocity sensor
Hide thumbs Also See for TIENet LaserFlow 360:
Table of Contents

Advertisement

Figure 4-12 LaserFlow placement using Slope Percentage
option
H - Optical Clarity
2%
slope
Optical Clarity – Sense voltage detects fog on the window of the
LaserFlow sensor. When the sense voltage is greater than 100
mV, the defogger is activated. The defogger is enabled from 1 to 5
minutes, depending on the severity of the fog.
Note
In locations with severe fogging potential, Teledyne Isco rec-
ommends increasing battery capacity or monitoring the site
more frequently.
To enable Optical Clarity, click "Active". To disable Optical
Clarity, click "Passive". The default is "Active".
TIENet™ 360 LaserFlow™ Velocity Sensor
Section 4 Setup with 2160 LaserFlow Module
4-13