Teledyne TIENet LaserFlow 360 Installation And Operation Manual page 135

Velocity sensor
Hide thumbs Also See for TIENet LaserFlow 360:
Table of Contents

Advertisement

Compliance Statements
DECLARATION OF CONFORMITY
Standards to which Conformity is Declared:
Standard
EN61000-4-2:2008
EN61000-4-3:2006
/A1:2007 /A2:2010
EN61000-4-4:2004
/A1:2010
EN61000-4-5:2005
EN61000-4-6:2008
EN61000-4-11:2004
CISPR11/
EN 55011:2009
/A1:1020
EN61000-3-2:2005
/A1:2008 /A2:2009
EN61000-3-3:2008
We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and
Standards as of September 21. 2012.
USA Representative
-
_______________________________
Vikas V. Padhye Ph, D
Vice President and General Manager
4700 Superior Street
Lincoln, Nebraska 68504
Phone: 402-464-0231
Fax:
60-4362-017
Rev. A
Application of Council Directive:
Manufacturer's Name:
Manufacturer's Address:
Equipment Type/Environment:
Trade Name/Model No:
Year of Issue:
Description
Electrostatic Discharge
Radiated RF Immunity
Electrical Fast Transient (EFT) on
Mains and I/O
Surge on AC Lines
Conducted RF Immunity on Mains and
I/O
Voltage Dips
RF Emissions Radiated, below 1GHz
and Conducted, AC Mains
AC Harmonics, Flicker
402-464-0318
2004/108/EC -The EMC Directive
2012/19/EC– The WEEE Directive
2006/95/EC – The Low Voltage Directive
Teledyne Isco
4700 Superior, Lincoln, Nebraska 68504 USA
Mailing Address: P.O. Box 82531, Lincoln, NE 68501
Phone: +1 (402) 464-0231
FAX:
+1 (402) 465-3799
Laboratory Equipment for Light Industrial/Commercial Environments
360 Laser Flow Sensor with 2160 Laser Flow Area Velocity Module and 2191
Battery Module
2012
EN 61326:2006 EMC Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use
nd
EN 61010-1 2
edition Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
EN60529 Special Protection offered by the Signature's Enclosure: IP-66
Severity Applied
Level 2 - 4kV contact discharge
Level 3 - 8kV air discharge
80 mHz to 2.7gHz 80% AM at 1 kHz
Level 2 - 3V/m
Level 2 - 1kV on AC lines
Level 2 - 1kV Line Common Mode
Level 2 - 0.5kV Differential Mode
150 kHz to 80 mHz
Level 1 – 1V rms, 80% Modulated
0% during1 cycle and full cycle
70% at 25 cycles
Group 1, Class A Industrial, Scientific, and
Medical Equipment
Performance Criteria
A
A
A
A
A
A
PASS
PASS

Advertisement