TIENet™ 360 LaserFlow™ Velocity Sensor
Section 2 Preparation and Installation
2.6.1 Wall Mount
Installation
1. Bolt or anchor the bottom
angle bracket to the wall.
2. Slide the mounting shaft
into the bottom angle
bracket and bolt or anchor
the mounting shaft to the
wall.
3. Insert either the long or
short tube coupler into the
hole formed by the vertical
clamp. Adjust the clamps to
the desired height and
cross-tighten the four bolts
evenly until the clamp and
tube coupler are firmly
attached.
Figure 2-19 Wall Mount: Vertical clamp and vertical clamp with tube coupler
inserted
2-18
Refer to the dimensions and requirements in Figure 2-17 while
performing the following steps.
Note
Stud anchors for concrete are provided in the mounting kit. It is
recommended to drill a hole 2" (5.08 cm) deep, leaving 1"
(2.54 cm) of the stud exposed.
Bottom Angle
Bracket
Mounting
Shaft
Tube
Coupler
Vertical
Clamp