TIENet™ 360 LaserFlow™ Velocity Sensor
Section 3 Setup with Signature® Flow Meter
carrier as necessary. By default, this option is enabled when the
"Enable Slope Setting" box is selected. See Figure 4-11 in Section
4.
Slope Percentage – If the slope of the pipe is known, mount
the sensor level with the Earth, using the bubble level mounted
on the top of the sensor and then enter the Slope Percentage in
the field. By default, this option is disabled when the "Enable
Slope Setting" box is selected. See Figure 4-12 in Section 4.
Optical Clarity – Sense voltage detects fog on the window of the
LaserFlow sensor. When the sense voltage is greater than 100
mV, the defogger is activated. The defogger is enabled from 1 to 5
minutes, depending on the severity of the fog.
Note
In locations with severe fogging potential, Teledyne Isco rec-
ommends increasing battery capacity or monitoring the site
more frequently.
To enable Optical Clarity, select "Active". To disable Optical
Clarity, select "Passive". The default is "Active".