Temporary Mounting Installation; Temporary Mount: Remove Stationary Foot; Temporary Mount: Install Coupler, Replace Stationary Foot - Teledyne TIENet LaserFlow 360 Installation And Operation Manual

Velocity sensor
Hide thumbs Also See for TIENet LaserFlow 360:
Table of Contents

Advertisement

2.7.1 Temporary Mounting
Installation
Mobile
End
Figure 2-24 Temporary Mount: Remove stationary foot
Figure 2-25 Temporary Mount: Install coupler, replace stationary foot
The temporary mounting kit is rated for up to 45 lbs (20.5 kg).
When installing the sensor with the temporary mounting kit,
ensure that the sensor carrier is positioned for easy installation
and removal of the sensor.
1. Slide the foot out of the stationary end of the cargo bar.
Release
Pump
Button
Handle
2. Slide the coupler onto the bar, tightening the two coupler
bolts (see Figure 2-26) enough to prevent it from sliding
around during the rest of the installation.
3. Reinsert the foot in the stationary end of the cargo bar.
4. In the manhole or vault, position the cargo bar for installa-
tion. Place the mobile end against the wall. and extend the
stationary end as far as possible.
5. Pump the handle fully up and down approximately 10
times, until the hydraulic tension presses the stationary
foot firmly against the opposite wall.
Note
To release the hydraulic pressure, press the red rubber button
next to the pump handle.
TIENet™ 360 LaserFlow™ Velocity Sensor
Section 2 Preparation and Installation
Stationary
End
2-23