Disassembly Guide

Front Case Assembly Replacement

54
11. Perform the performance verification tests before placing the monitor into
service. See Pulse Oximetry Functional Tests on page 28.
The front case assembly contains the keypad.
To replace the front case assembly:
1. Remove the batteries from the monitor. See Prior to Disassembly on page 48.
2. Separate the front and rear case assemblies. See Separating the Front and
Back Cases on page 48.
3. Dispose of the front case assembly in accordance with local regulations.
4. Connect the new front case assembly ribbon cable to the SpO
J4. See Figure 14.
Figure 14: Front Case Ribbon Cable Connection
1 — J4
O
M
XI
PCB connector
2
N-65 Service Manual