Performance Verification
30
Table 3 shows the Pulse Oximetry Functional Tests you must perform and a
description of each.
Table 3: Pulse Oximetry Functional Tests
Test
Description
Test Setup
This procedure establishes the baseline for all the other
tests.
Test Setup must be performed before performing any or
all of the SRC-MAX tests.
Test 1: BPM
This procedure simulates an OXIMAX sensor attached to a
patient indicating 60 bpm and 200 bpm.
The test sets up Test 1 for 60 bpm.
See Test 1: BPM on page 32.
Test 2: SpO
This procedure simulates an OXIMAX sensor attached to a
2
patient, indicating 75 percent blood oxygen saturation and
90 percent blood oxygen saturation.
The test sets up Test 2 for 75 percent blood oxygen
saturation.
See Test 2: SpO2 on page 33.
Test 3: Modulation
This procedure simulates an OXIMAX sensor attached to a
patient indicating low and high pulse strength.
The test sets up Test 3 for low pulse strength.
See Test 3: Modulation Level on page 34.
Test 4: Light
This procedure simulates an OXIMAX sensor attached to a
patient indicating low and high light level passing through
the patient at the sensor site.
The test sets up Test 4 for low light level.
See Test 4: Light Level on page 36.
See Test Setup on page 31.