Test 2: SpO
2
O
M
N-65 Service Manual
XI
AX
2. Press the SRC-MAX Pulse Rate select button.
The SRC-MAX Pulse Rate 60 LED lights.
The monitor pulse rate decreases to 60 and stabilize at 60 bpm. The test pass
criteria is 57 to 63 bpm inclusive.
The monitor:
displays 75 %SpO
displays 60 bpm (test pass criteria is 57 to 63 bpm inclusive)
alarms
Pulse Amplitude indicator displays low level modulation
To perform Test 2: SpO
1. Press the SRC-MAX %SpO2 select button.
The SRC-MAX %SpO
The monitor displays three dashes until the SRC-MAX stabilizes at 90
%SpO
. The test pass criteria is 88 to 92 %SpO
2
The monitor:
displays 90 %SpO
displays 60 bpm
no alarms
Pulse Amplitude indicator displays low level modulation
2
:
2
90 LED lights.
2
(pass criteria is 88 to 92 %SpO
2
Performance Verification
inclusive.
2
inclusive)
2
33