Safety Tests

O
M
N-65 Service Manual
XI
AX
4. Press the SRC-MAX Light Level selection button.
The SRC-MAX Light Level
The monitor pulse blip bar initially increases in amplitude and then reduces
and stabilizes at 3-5 bars.
The monitor:
displays 75 %SpO
displays 60 bpm
alarms
pulse amplitude indicator displays low level modulation
5. Perform Test 1: BPM on page 32.
The pulse amplitude indicator should indicate low level modulation.
6. Perform Test 2: SpO2 on page 33.
The pulse amplitude indicator should indicate low level modulation.
7. Disconnect all equipment and turn off the monitor.
The monitor safety tests meet the standards of, and are performed in accordance
with, EN 60601-1: 1990 (A1 and A2), IEC 60601-1, UL 60601-1, and CAN/CSA
C22.2 No. 601.1. This product is categorized as internally powered type BF
equipment.
LED lights.
2
Performance Verification
37