5. Lock the ribbon cable connector. See Figure 15.
Figure 15: Locking SpO
6. Place the front and rear cases together and install the four screws. See
Figure 16, items 1 through 4.
Figure 16: Installing The Case Screws
1 — SpO
PCB
2
2 — J4 lock
3 — Lock direction
4 — Unlock direction
PCB Connection to J4