Specifications
70
Audible Indicators
Audible
Parameter
Indicator
Alarm Volume
Volume level
Setting
Pitch (±30 Hz)
On pulse width
(±20 msec)
Off Interval (±20 msec)
Number of pulses in burst 1
Repetition Pause
(±2 sec.)
Priority
Beep Volume
Volume level
setting
Pitch (±30 Hz)
On pulse width
(±20 msec)
Off Interval (±20 msec)
Number of pulses in burst 1
Repetition Pause
(±2 sec.)
Priority
POST Pass
Volume level
Pitch (±30 Hz)
On pulse width
(±20 msec)
Off Interval (±20 msec)
Number of pulses in burst 1
Repetition Pause
(±2 sec.)
Priority
Invalid Key Press
Volume level
Pitch (±30 Hz)
On pulse width
(±20 msec)
Off Interval (±20 msec)
Number of pulses in burst 1
Repetition Pause
(±2 sec.)
Priority
Value
Adjustable, 40 to 52
dB(A), at one meter
752 Hz
500 msec
10 msec
N/A
1
Adjustable, 42 to 52
dB(A), at one meter
1500 Hz
500 msec
10 msec
N/A
2
Fixed at 45 dB(A), at one
meter
600 Hz
1000 msec
10 msec
N/A
3
Fixed at 45 dB(A), at one
meter
200 Hz
50 msec
10 msec
N/A
4