M
N-65 Service Manual
XI
AX
2. Perform Test 1: BPM on page 32.
The Pulse Amplitude indicator should indicate high level modulation.
3. Perform Test 2: SpO2 on page 33.
The Pulse Amplitude indicator should indicate high level modulation.
4. Press the SRC-MAX % Modulation selection button.
The SRC-MAX % Modulation
The monitor pulse blip bar decreases in amplitude.
The monitor:
•
displays 75 %SpO
•
displays 60 bpm
•
alarms
•
Pulse Amplitude indicator displays low level modulation
5. Perform Test 1: BPM on page 32.
The pulse amplitude indicator should indicate low level modulation.
6. Perform Test 2: SpO2 on page 33.
The pulse amplitude indicator should indicate low level modulation.
LED lights.