Nellcor OxiMax N-65 Service Manual page 50

Handheld
Hide thumbs Also See for OxiMax N-65:
Table of Contents

Advertisement

Troubleshooting
Scope of Action
Error
Code
257
258
259
260
261
262
263
264
265
266
267
268
269
270
271
272
273
1, 2
274
275
X
276
X
277
X
280
X
281
X
282
X
44
Table 5: Error Codes (Continued)
Action
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
X
Replace SpO
3
1 — Restart the monitor.
2 — Set the time and date.
3 — Replace UI board.
X
Return monitor for
reprogramming.
Check/replace sensor/
extension cable.
Replace with O
extension cable.
Check/replace sensor/
extension cable.
Check/replace sensor/
extension cable.
Check/replace sensor/
extension cable.
Check/replace sensor/
extension cable.
Explanation
board.
SpO
back end reports packet start ID
2
2
(SID) missing.
board.
SpO
back end reports packet length
2
2
error.
board.
SpO
back end reports message length
2
2
error.
board.
SpO
back end reports packet contains
2
2
unsupported key.
board.
SpO
back end reports packet CRC
2
2
error.
board.
SpO
back end reports end of packet
2
2
missing.
board.
SpO
back end reports packet contains
2
2
unidentified key.
board.
SpO
back end reports corrupted
2
2
variable.
board.
SpO
back end reports memory
2
2
overflow.
board.
SpO
back end reports bad pointer.
2
2
board.
SpO
back end reports parameter
2
2
value out-of-range.
board.
SpO
back end reports reset detected.
2
2
board.
SpO
back end reports unexpected
2
2
value.
board.
SpO
back end reports time-out.
2
2
board.
SpO
back end reports not ready/not
2
2
initialized.
board.
SpO
back end reports double fault.
2
2
SpO
back end reports date out-of-
2
range error.
SpO
back end reports incompatible
2
software version.
SpO
back end reports incorrect
2
registration number.
M
sensor/
SpO
back end reports sensor read
XI
AX
2
failure.
SpO
back end reports sensor
2
signature verification fails.
SpO
back end reports does not
2
support feature required by sensor.
SpO
back end reports overflow/
2
underflow.
SpO
back end reports sensor
2
activation failure.
O
M
N-65 Service Manual
XI
AX

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents