3.6 Connecting the Compressor to the
Concentrator

3.6.1 Connecting the Compressor to Invacare

Concentrators

1. Insert one end of the interconnect hose C into the
outlet fitting B on the back of the concentrator A.
2. Insert the opposite end of the interconnect hose into
the inlet fitting D on the side of the compressor.
The metal tabs E on the concentrator outlet
fitting and the compressor inlet fitting will pop
out with an audible "click" when the end of the
interconnect hose is properly installed.
1195663-C

3.6.2 Connecting the Compressor to Non-Invacare

Concentrators
1. Refer to the oxygen concentrator user manual outlet
gas specifications to verify compatibility with the
HomeFill compressor input gas requirements listed in
8.3 Specifications, page 35.
2. Refer to the oxygen concentrator user manual for
connecting instructions.