Cartridge Instructions For Use; Filling And Loading A T:slim Cartridge - Tandem T:Slim User Manual

Hide thumbs Also See for T:Slim:
Table of Contents

Advertisement

Chapter 5 – Infusion Site Care and Loading Cartridge
5.2
Cartridge Instructions
for Use
For complete cartridge labeling, consult
the Cartridge Instructions for Use
included in the t:slim Cartridge box.
5.3
Filling and Loading a
t:slim Cartridge
This section describes how to fill the
cartridge with insulin and load the
cartridge into your t:slim Pump. The
single-use disposable cartridge can
hold up to 300 units (3.0 mL) of insulin.
PRECAUTION
Replace the cartridge every 48 hours if using
Humalog
; every 72 hours if using Novolog
®
PRECAUTION
Inspect the cartridge to ensure that there is no
damage. Never use a cartridge that is damaged
(cracks, dislodged tip connectors, etc.). If
there is damage, call Tandem Diabetes Care
Customer Technical Support at 1-877-801-
6901 to request a replacement cartridge.
64
PRECAUTION
Before you begin, verify that the infusion set is
NOT connected to your body.
Before you begin, make sure you
have the following items:
1 unopened cartridge
3.0 mL syringe and fill needle
Vial of Humalog
®
insulin
Alcohol prep swab
1 new infusion set
Infusion set Instructions for Use
NOTE
.
®
Do NOT remove the used cartridge from the
pump during the load process until prompted
on the t:slim Pump screen.
The illustration identifies the
luer-lock and insulin fill port used in
the cartridge filling process.
Insulin
Fill Port
or NovoLog
®
Customer Technical Support: 1-877-801-6901
Luer-Lock
Tail of the
Cartridge

Hide quick links: