Chapter 19 – Technical Specifications

19.1 Overview

This appendix provides tables of technical specifications, options, settings, and electromagnetic compliance information for
the t:slim System.

19.2 t:slim Pump Specifications

The following table lists the t:slim Pump specifications and their related details.
Specification Type
Classification
Size
Weight (with full disposable)
Operating Temperature Range
Storage Temperature Range
Atmospheric Pressure
Reservoir Volume
Cannula Fill Amount
Insulin Concentration
Alarm Type
table continued on next page ...
166
Specification Details
External PSU: Class II, Infusion Pump. Internally-powered equipment, Type BF applied part. The risk of ignition
of flammable anesthetics and explosive gases by the pump is remote. While this risk is remote, it is not
recommended to operate the t:slim Pump in the presence of flammable anesthetics or explosive gases.
3.13˝ x 2.0˝ x 0.6˝ (L x W x H) – (7.95 cm x 5.08 cm x 1.52 cm)
3.95 ounces (112 grams)
41°F and 98.6°F, 5°C and 37°C at maximum relative humidity of 90% non-condensing
-4°F and 140°F -20°C to 60°C at maximum relative humidity of 90% non-condensing
700 hPa to 1060 hPa (equivalent from 10,000 ft. to Dead Sea elevation)
3.0 mL or 300 units
0.1 to 1.0 units of insulin
U-100
Visual, audible, and vibratory
Customer Technical Support: 1-877-801-6901