Chapter 16 – t:slim Pump Alarms
16.5 Cartridge Removal Alarm
What will I see on the Screen?
CARTRIDGE ALARM
ALL DELIVERIES STOPPED!
The cartridge cannot be detected.
Press INSTALL to install a new
cartridge or press CONNECT to
reconnect the current cartridge.
CONNECT
INSTALL
150
What does it mean?
Your t:slim Pump detected that the cartridge has been removed and all deliveries
have stopped.
How will the system notify me?
3 sequences of 3 notes or 3 vibrations depending on the volume/vibrate setting
selected in Pump Volume.
Will the system re-notify me?
Yes. The system will re-notify you every 3 minutes until you reconnect the current
cartridge or change the cartridge.
How should I respond?
Tap CONNECT to reattach the current cartridge. Tap INSTALL to load a
new cartridge.
Customer Technical Support: 1-877-801-6901