Chapter 19 – Technical Specifications

19.6 Electromagnetic Immunity

The t:slim System is intended for use in the electromagnetic environment specified below. Always make sure that the System is
used in such an environment.
Immunity Test
IEC 60601 Test Level
Electrostatic Discharge (ESD)
± 8 kV contact
IEC 61000-4-2
± 15 kV air
Electrical Fast Transient/burst
± 2 kV for power supply lines
IEC 61000-4-4
± 1 kV for input/ output lines
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
Voltage Dips, Short Interruptions,
<5% Ur (>95% dip in Ur) for 0,5 cycle
and Voltage Variations on Power
40% Ur (60% dip in Ur) for 5 cycles
Supply Input Lines IEC 61000-4-11
70% Ur (30% dip in Ur) for 25 cycles
<5% Ur (>95% dip in Ur) for 5 sec
Power Frequency (50/60 Hz)
3 A/m
Magnetic Field IEC 61000-4-8
table continued on next page ...
176
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Compliance Level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
<5% Ur (>95% dip in Ur) for 0,5 cycle
40% Ur (60% dip in Ur) for 5 cycles
70% Ur (30% dip in Ur) for 25 cycles
<5% Ur (>95% dip in Ur) for 5 sec
400 A/m (IEC 60601-2-24)t
Electromagnetic Environment –
Guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the pump requires
continued operation during power mains interruptions,
it is recommended that the pump be powered from an
uninterruptible power supply or a battery.
NOTE: Ur is the a.c. mains voltage prior to application of
the test level.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.