Tandem T:Slim User Manual page 137

Hide thumbs Also See for T:Slim:
Table of Contents

Advertisement

Max Basal Alert 2
What will I see on the Screen?
MAX BASAL ALERT
You have exceeded 2x your highest
basal setting. Please review your
current temp rate in the Options
menu.
CLOSE
Customer Technical Support: 1-877-801-6901
What does it mean?
An active temp rate exceeded 2 times your highest basal setting defined in your
Personal Profile.
How will the system notify me?
2 sequences of 3 notes or 2 vibrations depending on the volume/vibrate setting
selected in Pump Volume.
Will the system re-notify me?
Yes, every 5 minutes until acknowledged.
How should I respond?
Tap CLOSE and review your current temp rate in the Options menu.
Chapter 15 – t:slim Pump Alerts
137

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents

Save PDF