Tandem T:Slim User Manual page 167

Hide thumbs Also See for T:Slim:
Table of Contents

Advertisement

...table continued from previous page
Specification Type
Basal Delivery Accuracy at all Flow
Rates (tested per IEC 60601-2-24)
Bolus Delivery Accuracy at all Volumes
(tested per IEC 60601-2-24)
Patient Protection from Air Infusion
Maximum Infusion Pressure Generated
and Occlusion Alarm Threshold
Rate of delivery during Bolus
Rate of delivery during Prime
Bolus Duration
Frequency of Basal Delivery
Retention Time of Electronic Memory
when Internal System Battery is
Fully Discharged (including Alarm
Settings and Alarm History)
Infusion Set used for Testing
table continued on next page ...
* In certain pressure environments, such as a gradual elevation change of 1,000 feet in 15 minutes, delivery accuracy may vary ±15% for short periods of time.
Customer Technical Support: 1-877-801-6901
Specification Details
±5% *
±5% *
The pump provides subcutaneous delivery into interstitial tissue and does not deliver intravenous injections.
Clear tubing aids in detecting air.
30 PSI
Bolus > 3 units
Approximately 2.8 units/min
Bolus ≤ 3 units
Approximately 1.6 units/min
Approximately 9.7 units/min
Bolus > 3 units duration
Approximately 21 seconds/unit
Bolus ≤ 3 units duration
Approximately 38 seconds/unit
5 minutes for all Basal Rates
Greater than 30 days
Unomedical Comfort
Infusion Set
Chapter 19 – Technical Specifications
167