16.6 Temperature Alarm

What will I see on the Screen?
TEMPERATURE ALARM
ALL DELIVERIES STOPPED!
Remove pump from extreme
temperatures and then resume
insulin delivery.
CLOSE
Customer Technical Support: 1-877-801-6901
What does it mean?
Your t:slim Pump detected a temperature below 35.6 °F (2 °C) or above 113 °F
(45 °C) and all deliveries have stopped.
How will the system notify me?
3 sequences of 3 notes or 3 vibrations depending on the volume/vibrate setting
selected in Pump Volume.
Will the system re-notify me?
Yes. The system will re-notify you every 3 minutes until a temperature in the operating
range is detected.
How should I respond?
Tap CLOSE. Remove the pump from the extreme temperature and then resume
insulin delivery.