16.9 Altitude Alarm

What will I see on the Screen?
ALTITUDE ALARM
ALL DELIVERIES STOPPED!
Remove cartridge from pump,
reconnect cartridge and then
resume insulin.
CLOSE
Customer Technical Support: 1-877-801-6901
What does it mean?
Your t:slim Pump detected a pressure change outside of the operating range and all
deliveries have stopped.
How will the system notify me?
3 sequences of 3 notes or 3 vibrations depending on the volume/vibrate setting
selected in Pump Volume.
Will the system re-notify me?
Yes. The system will re-notify you every 3 minutes until the condition is corrected.
How should I respond?
Tap CLOSE. Remove the cartridge from the pump (this will allow the cartridge to fully
vent) and then reconnect the cartridge.