Chapter 15 – t:slim Pump Alerts

15.12 Connection Error Alert

What will I see on the Screen?
CONNECTION ERROR ALERT
Pump cannot connect with the
computer. Press CLOSE and
reconnect the USB cable to try
again.
CLOSE
140
What does it mean?
You connected your t:slim Pump to a computer with the USB cable to charge it or
upload data to t:connect and a connection could not be made.
How will the system notify me?
2 sequences of 3 notes or 2 vibrations depending on the volume/vibrate setting
selected in Pump Settings.
Will the system re-notify me?
Yes, every 5 minutes until acknowledged.
How should I respond?
Tap CLOSE. Disconnect and reconnect the USB cable to try again.