T:slim Pump Options And Settings - Tandem T:Slim User Manual

Hide thumbs Also See for T:Slim:
Table of Contents

Advertisement

Chapter 19 – Technical Specifications

19.3 t:slim Pump Options and Settings

The following table lists the t:slim Pump options and settings, and their related details.
Option/Setting Type
Time
Maximum Basal Rate
Insulin Delivery Profiles
(Basal and Bolus)
Basal Rate Segments
Basal Rate Increment
Temp Basal Rate
Bolus Setup
Insulin-to-Carb (IC) Ratio
BG Correction Target Value
Insulin Sensitivity Factor (ISF)
Duration of Insulin Action
Bolus Increment
table continued on next page ...
172
Option/Setting Details
12-hour clock
15 units/hr
6
16 per delivery profile
0.001 at programed rates equal to or greater than 0.1 units/hr
15 minutes to 72 hours with 1 minute resolution with a range of 0% to 250%
Can deliver based on carb input (grams) or insulin input (units). Default is units.
(The range for carbs is 1–999 grams; the range for insulin is 0.05–25 units)
16 time segments per 24-hour period; Ratio: 1 unit of insulin per x grams of carbs; 1:1 to 1:300
(can be set by 0.1 below 10)
16 time segments. 70 to 250 mg/dL in 1 mg/dL increments
16 time segments; Ratio: 1 unit of insulin reduces glucose x mg/dL; 1:1 to 1:600
(1 mg/dL increments)
1 time segment; 2–8 hours in 1-minute increments (default is 5 hrs)
0.01 at volumes greater than 0.05 units
Customer Technical Support: 1-877-801-6901

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents