Troubleshooting; Troubleshooting Chart - CAS LP-II Service Manual

Hide thumbs Also See for LP-II:
Table of Contents

Advertisement

16 Troubleshooting

16.1 Troub leshooting Chart
Symptom
Possible Causes
Scale will not
1. No power to the Power
power ON.
Supply.
2. No power to Main PCB.
3. Power to Main PCB is OK.
Scale
1. "Printer Continuously Busy."
powers ON
with a n error.
2. "Calibrated mass error."
3. "Load Cell not ready."
4. "Zero Range Error."
For more than 45 seconds.
5. Scale beeps and "looks"
like it is turned ON, but there
is no display light.
Scale
1. Scale weight is inaccurate.
p
owers ON
&
goes to
Z
ero.
Possible Solutions
Plug the Power Cord in to a working outlet
Check/replace fuse.
Check/replace Power Cord.
Check/replace Main Power Switch.
Check/replace Power Supply Inpu
Check/replace Wire Harness, CN21 PS.
Check/replace Wire Harness, CN22 PS.
Check/replace Power Supply.
Check/replace Display Harness, J17/J18 M
Check/replace Display Stalk.
Check/replace ADC Wire H
Check/replace AD-PCB.
Check/replace Ne
twork PCB.
Check/replace Printer Assembly.
Check/replace all EPROMs.
Check/replace Memory Expansion P
Check/replace Main PCB.
Also check 2.No power to Main PCB.
Check/replace P
rinter System EPROM, U8 M-PCB.
Check/replace Main PCB.
Check/replace Printer Assembly.
Reprogram all Weight settings & Span ca
scale.
Check for platter or platform obstruction.
Check/replace ADC Wire Harness, CON3 AD-PCB.
Check/replace ADC Wire Harness, CON2 AD-PCB.
Check/replace AD-PCB.
Check/replace Load Cell Assembly.
Check for platter or platform obstruction.
Remove all items from platter and cycle power.
Reprogram all Weight settings & Span calibrate the
scale.
Check/replace ADC Wire Harness, CON2 AD-PCB.
Check/replace AD-PCB.
Check/replace Load Cell Assembly.
Check/replace Main PCB.
Check/replace Display Harness, J17/J18 M-PCB.
Check/replace Display Stalk.
Check/replace Wire Harness, CN21 PS.
Check/replace Power Supply.
Check/replace Main PCB.
Check for platter or platform obstruction.
Remove all items from platter and cycle power.
Reprogram all Weight settings & Span calibrate the
scale.
Check 3."Load Cell Not Ready."
57
.
t Harness, CN1 PS.
-PCB.
arness, CON3 AD-PCB.
CB.
librate the
Reference
3.3, 4.1,
14.4, 17.5,
18.1, 19.7
2.1, 2.2, 3
.3,
4.1, 4.2,
14.7 14.3,
14.4 14.6
17.4, 17.
5,
18.1, 19.3,
19.6, 19.7
8.12, 12.10,
13.2, 13.3,
1
7.3, 17.5,
1
9.1, 19.6
10.5, 12.10,
14.3, 17.3,
19.6
7.1 ~ 7.4,
6.1
2.1, 4.1, 6.1,
7.1 ~ 7.4,
12.3, 13.2,
13.5, 14.5,
19.5
2.1, 4.1, 6.1,
7.1 ~ 7.4,
12.3,
14.1 ~ 14.5
17.4, 19.5
2.1, 2.2, 4.1,
4.2, 11.2,
14.3, 14.4,
14.7, 17.3,
19.3, 19.6
2.1, 4.1, 6.1,
7.1 ~ 7.4,
12.3, 14.3,
14.5, 17.4,
19.5

Hide quick links: