3 Proper Operation
3.2

Leveling and Location

1) Location:
This scale must be placed on a flat and stable surface. Please keep the scale away from
the direct path of oscillating fans, ventilation systems, or strong drafts as these air disturbances can be
picked-up by the scale's very sensitive weighing platform and may cause incorrect weight readings.
2) Leveling:
If the scale is not properly leveled, please adjust the 4 adjustable legs at the bottom of
the scale. Turn the legs clockwise or counterclockwise so as to center the bubble of the leveling gauge
inside the indicated circle. Turning the adjustable legs counter-clockwise (viewed from top of scale) will
lower that part of the scale. Turning the adjustable legs clockwise (viewed from top of scale) will raise
that part of the scale. (See Fig.)