Survey of Reference and max Hearing Level Speech Audiometer.

ANSI Speech.
ANSI
Coupler: ANSI S3.7-1995 (NBS-9A) / IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
ANSI S3.6-2010
RETSPL
Linear Type A
19.5
Equ. FF Type A-E
15.5
Not Linear¹
7.0
ILTASS
ANSI
Coupler: ANSI S3.7-1995 (NBS-9A) / IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
RETSPL
Linear Type A
20.0
Equ. FF Type A-E
21.5
Not Linear¹
7.5
ILTASS
ANSI
HDA 200
Coupler: IEC60318-2 with type 1 adaptor
Speech Audiometer
Speech
ANSI S3.6-2010
Frequency
RETSPL MaxHL RETSPL MaxHL
Linear Type A
19.0
Equ. FF Type A-E
18.5
Not Linear¹
5.5
ILTASS
ANSI
Coupler: ANSI S3.7-1995 (NBS-9A) / IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
ANSI S3.6-2010
RETSPL
Linear Type A
18.5
Equ. FF Type A-E
18.5
Not Linear¹
6.0
ILTASS
Binaural
45°
RETSPL RETSPL RETSPL RETSPL RETSPL RETSPL
Speech
15.0
11.0
Speech Noise
15.0
11.0
ILTASS
15.0
11.0
TDH39
Speech Noise
ANSI S3.6-2010
MaxHL
RETSPL
MaxHL
110
19.5
100
105
15.5
100
120
7.0
115
19.5
105
HDA 280
Speech Noise
MaxHL
RETSPL
MaxHL
100
20.0
95
21.5
120
7.5
120
20.0
100
Speech Noise
ANSI S3.6-2010
90
19.0
85
85
18.5
80
110
5.5
105
19.0
90
DD45
Speech Noise
ANSI S3.6-2010
MaxHL
RETSPL
MaxHL
110
18.5
100
100
18.5
100
120
6.0
115
18.5
105
ANSI
Free Field
ANSI S3.6-2010
Monaural
90°
9.5
17.0
9.5
17.0
9.5
17.0
¹ This value come from pure tone 1kHz RETSPL value. (Not linear
describe there is no equalizing of the signal flat electrical response)
Coupler: ANSI S3.7-1995 (HA-2 with 5mm rigid Tube)
Speech
ILTASS
95
90
Speech
ILTASS
Coupler: ANSI S3.7-1995 (HA-2)
Speech
ILTASS
Coupler: ANSI S3.13-1987 (R2002)
Speech
ILTASS
Free Field max HL is found by subtracting the selected RETSPL value
Free Field Power
45°
90°
0° - 45° - 90°
Max SPL
13.0
11.5
90
13.0
11.5
85
13.0
11.5
90
ANSI
Ear 3A
Speech Audiometer
Speech
Speech Noise
ANSI S3.6-2010
ANSI S3.6-2010
RETSPL
MaxHL
RETSPL
12.5
110
12.5
12.5
ANSI
Ear 5A
Coupler: IEC 60318-5 2006
Speech Audiometer
Speech
Speech Noise
ANSI S3.6-2010
ANSI S3.6-2010
RETSPL MaxHL RETSPL MaxHL
12.5
110
12.5
12.5
ANSI
CIR 22
Speech Audiometer
Speech
Speech Noise
ANSI S3.6-2010
ANSI S3.6-2010
RETSPL MaxHL RETSPL MaxHL
12.5
110
12.5
12.5
ANSI
B71
Speech Audiometer
Speech
Speech Noise
ANSI S3.6-2010
ANSI S3.6-2010
RETSPL MaxHL RETSPL MaxHL
55.0
50
55.0
55.0
Free Field max SPL
Free Field Line
0° - 45° - 90°
Max SPL
100
100
100
MaxHL
100
100
100
100
100
100
50
50
Free Field Speaker
0° - 45° - 90°
Max SPL
80
75
75