Interacoustics AD629 Instructions For Use Manual page 191

Diagnostic audiometer
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Sweden Speech Calibration
SWEDEN
Coupler IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
STAF 1996
RETSPL MaxHL RETSPL MaxHL
Linear Type A
22.0
Equ. FF Type A-E
0.5
Not Linear
22.0
ILTASS
SWEDEN
Coupler IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
IEC60645-2 1997
RETSPL MaxHL RETSPL MaxHL
Linear Type A
20.0
Equ. FF Type A-E
6.5
Not Linear
7.5
ILTASS
SWEDEN
Coupler IEC60318-1 with type 1 adaptor
Speech Audiometer
Speech
IEC60645-2 1997
Frequency
RETSPL MaxHL RETSPL MaxHL
Linear Type A
20.0
Equ. FF Type A-E
3.5
Not Linear
5.5
ILTASS
SWEDEN
Coupler IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
STAF 1996
RETSPL MaxHL RETSPL MaxHL
Linear Type A
22.0
Equ. FF Type A-E
3.5
Not Linear
22.0
ILTASS
Binaural
RETSPL RETSPL RETSPL RETSPL RETSPL RETSPL
Speech
0.0
Speech Noise
0.0
ILTASS
0.0
TDH39
Speech Noise
STAF 1996
105
27.0
90
120
0.5
115
105
27.0
90
27.0
90
HDA 280
Speech Noise
IEC60645-2 1997
100
20.0
95
110
6.5
105
120
7.5
120
20.0
100
HDA 200
Speech Noise
IEC60645-2 1997
90
20.0
85
100
3.5
95
110
5.5
105
20.0
90
DD45
Speech Noise
STAF 1996
105
27.0
90
115
3.5
115
105
27.0
90
27.0
90
Sweden Free Field
ISO 389-7 2005
45°
90°
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
Speech
ILTASS
Speech
ILTASS
Speech
ILTASS
Speech
ILTASS
Free Field max HL is found by subtracting the selected RETSPL value
Monaural
Free Field Power
45°
90°
0° - 45° - 90°
Max SPL
0.0
0.0
0.0
0.0
0.0
0.0
SWEDEN
EAR 3A
Coupler IEC 60318-5 2006
Speech Audiometer
Speech
Speech Noise
STAF 1996
STAF 1996
RETSPL MaxHL RETSPL MaxHL
21.0
100
26.0
26.0
SWEDEN
EAR 5A
Coupler IEC 60318-5 2006
Speech Audiometer
Speech
Speech Noise
STAF 1996
STAF 1996
RETSPL MaxHL RETSPL MaxHL
21.0
100
26.0
26.0
SWEDEN
CIR 22
Coupler IEC 60318-5 2006
Speech Audiometer
Speech
Speech Noise
STAF 1996
STAF 1996
RETSPL MaxHL RETSPL MaxHL
21.0
90
26.0
26.0
SWEDEN
B71
Coupler 60318-6 2007
Speech Audiometer
Speech
Speech Noise
IEC60645-2 1997
IEC60645-2 1997
RETSPL MaxHL RETSPL MaxHL
55.0
50
55.0
55.0
Free Field max SPL
Free Field Line
0° - 45° - 90°
Max SPL
90
100
85
100
90
100
85
85
85
85
85
85
50
50
Free Field Speaker
0° - 45° - 90°
Max SPL
80
75
75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents