Survey Of Reference And Max Hearing Level Speech Audiometer - Interacoustics AC40 Service Manual

Clinical audiometer
Table of Contents

Advertisement

AC40 Service Manual

Survey of reference and max hearing level Speech Audiometer

ANSI Speech
ANSI
Coupler: ANSI S3.7-1995 (NBS-9A) / IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
ANSI S3.6-2010
RETSPL
Linear Type A
19.5
Equ. FF Type A-E
15.5
Not Linear¹
7.0
Speech WN
ANSI
Coupler: ANSI S3.7-1995 (NBS-9A) / IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
RETSPL
Linear Type A
20.0
Equ. FF Type A-E
21.5
Not Linear¹
7.5
Speech WN
ANSI
HDA 200
Coupler: IEC60318-2 with type 1 adaptor
Speech Audiometer
Speech
ANSI S3.6-2010
Frequency
RETSPL MaxHL RETSPL MaxHL
Linear Type A
19.0
Equ. FF Type A-E
18.5
Not Linear¹
5.5
Speech WN
ANSI
Coupler: ANSI S3.7-1995 (NBS-9A) / IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech
ANSI S3.6-2010
RETSPL
Linear Type A
18.5
Equ. FF Type A-E
18.5
Not Linear¹
6.0
Speech WN
¹ This value come from pure tone 1kHz RETSPL value. (Not linear describe
there is no equalizing of the signal flat electrical response)
Binaural
RETSPL
RETSPL
Speech
15.0
Speech Noise
15.0
Speech WN
17.5
TDH39
Speech Noise
ANSI S3.6-2010
MaxHL
RETSPL
MaxHL
110
19.5
100
105
15.5
100
120
7.0
115
22.0
95
HDA 280
Speech Noise
MaxHL
RETSPL
MaxHL
100
20.0
95
95
21.5
90
120
7.5
120
22.5
95
Speech Noise
ANSI S3.6-2010
90
19.0
85
85
18.5
80
110
5.5
105
21.5
90
DD45
Speech Noise
ANSI S3.6-2010
MaxHL
RETSPL
MaxHL
110
18.5
100
100
18.5
100
120
6.0
115
21.0
95
ANSI
Free Field
ANSI S3.6-2010
45°
90°
RETSPL
RETSPL
11.0
9.5
17.0
11.0
9.5
17.0
13.5
12.0
19.5
Coupler: ANSI S3.7-1995 (HA-2 with 5mm rigid Tube)
Speech
Speech WN
Speech
Speech WN
Speech
Speech WN
Speech
Speech WN
Free Field max HL is found by subtracting the
Monaural
Free Field Power
45°
90°
0° - 45° - 90°
RETSPL
RETSPL
Max SPL
13.0
11.5
13.0
11.5
15.5
14.0
87.5
ANSI
Ear 3A
Speech Audiometer
Speech
Speech Noise
ANSI S3.6-2010
ANSI S3.6-2010
RETSPL MaxHL
RETSPL MaxHL
12.5
110
12.5
15.0
ANSI
Ear 5A
Coupler: IEC 60318-5 2006
Speech Audiometer
Speech
Speech Noise
ANSI S3.6-2010
ANSI S3.6-2010
RETSPL MaxHL RETSPL MaxHL
12.5
110
12.5
15.0
ANSI
CIR 22/CIR33
Coupler: ANSI S3.7-1995 (HA-2)
Speech Audiometer
Speech
Speech Noise
ANSI S3.6-2010
ANSI S3.6-2010
RETSPL MaxHL RETSPL MaxHL
12.5
100
12.5
15.0
ANSI
B71
Coupler: ANSI S3.13-1987 (R2002)
Speech Audiometer
Speech
Speech Noise
ANSI S3.6-2010
ANSI S3.6-2010
RETSPL MaxHL RETSPL MaxHL
55.0
60
55.0
57.5
Free Field max SPL
selected RETSPL value
Free Field Line
0° - 45° - 90°
Max SPL
90
100
85
100
97.5
Page 10
100
95
100
95
100
95
50
55

Hide quick links: