Survey Of Reference And Max Hearing Level Tone Audiometer - Interacoustics AD629 Instructions For Use Manual

Diagnostic audiometer
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Survey of Reference and max Hearing Level Tone Audiometer.

ANSI TDH39
Coupler: ANSI S3.7-1995 (NBS-9A) /
IEC 60318-3 1998 (6ccm)
Tone Audiometer
Tone
ANSI S3.6-2010
Frequency
RETSPL MaxHL
125
45.0
90
160¹
37.5
95
200¹
31.5
100
250
25.5
110
315¹
20.0
115
400¹
15.0
120
500
11.5
120
630¹
8.5
120
750
8.0
120
800¹
7.0
120
1000
7.0
120
1250¹
6.5
120
1500
6.5
120
1600¹
7.0
120
2000
9.0
120
2500¹
9.5
120
3000
10.0
120
3150¹
10.0
120
4000
9.5
120
5000¹
13.0
120
6000
15.5
120
6300¹
15.0
120
8000
13.0
110
WhiteNoise
TenNoise
¹ The RETSPL is copy from ISO389-1 1998
ANSI HDA 280
Coupler ANSI S3.7-1995 (NBS-9A) /
IEC 60318-3 1998 (6ccm)
Tone Audiometer
Tone
PTB
Frequency
RETSPL MaxHL
125
38.5
105
160¹
33.5
110
200¹
29.5
115
250
25.0
120
315¹
21.0
120
400¹
17.0
120
500
13.0
120
630¹
10.5
120
750
9.0
120
800¹
8.5
120
1000
7.5
120
1250¹
8.5
120
1500
9.5
120
1600¹
9.0
120
2000
8.0
120
2500¹
7.0
120
3000
6.5
120
3150¹
7.0
120
4000
9.5
120
5000¹
12.0
120
6000
19.0
115
6300¹
19.0
115
8000
18.0
105
WhiteNoise
¹ This value is interpolated on Interacoustics
Narrow Band Noise
ANSI S3.6-2010
RETSPL
MaxHL
49.0
75
41.5
85
35.5
90
29.5
95
24.0
100
19.0
105
15.5
110
13.5
110
13.0
110
12.0
110
13.0
110
12.5
110
12.5
110
13.0
110
15.0
110
15.5
110
16.0
110
16.0
110
14.5
110
18.0
110
20.5
110
20.0
110
18.0
100
0.0
120
25.0
110
Narrow Band Noise
PTB
RETSPL
MaxHL
42.5
75
37.5
80
33.5
85
29.0
90
25.0
95
21.0
95
17.0
100
15.5
100
14.0
105
13.5
105
13.5
105
14.5
105
15.5
105
15.0
105
14.0
105
13.0
105
12.5
105
13.0
105
14.5
105
17.0
105
24.0
95
24.0
95
23.0
90
0.0
120
IEC TDH39
Coupler: IEC 60318-3 1998 (6ccm)
Tone Audiometer
Tone
ISO 389-1 1998
Frequency
RETSPL MaxHL
125
45.0
90
160
37.5
95
200
31.5
100
250
25.5
110
315
20.0
115
400
15.0
120
500
11.5
120
630
8.5
120
750
7.5
120
800
7.0
120
1000
7.0
120
1250
6.5
120
1500
6.5
120
1600
7.0
120
2000
9.0
120
2500
9.5
120
3000
10.0
120
3150
10.0
120
4000
9.5
120
5000
13.0
120
6000
15.5
120
6300
15.0
120
8000
13.0
110
WhiteNoise
TenNoise
IEC HDA 280
Coupler IEC 60318-3 1998 (6ccm)
Tone Audiometer
Tone
PTB
Frequency
RETSPL MaxHL
125
38.5
105
160¹
33.5
110
200¹
29.5
115
250
25.0
120
315¹
21.0
120
400¹
17.0
120
500
13.0
120
630¹
10.5
120
750
9.0
120
800¹
8.5
120
1000
7.5
120
1250¹
8.5
120
1500
9.5
120
1600¹
9.0
120
2000
8.0
120
2500¹
7.0
120
3000
6.5
120
3150¹
7.0
120
4000
9.5
120
5000¹
12.0
120
6000
19.0
115
6300¹
19.0
115
8000
18.0
105
WhiteNoise
¹ This value is interpolated on Interacoustics
Narrow Band Noise
ISO 389-4 1994
RETSPL
MaxHL
49.0
75
41.5
85
35.5
90
29.5
95
24.0
100
19.0
105
15.5
110
13.5
110
12.5
110
12.0
110
13.0
110
12.5
110
12.5
110
13.0
110
15.0
110
15.5
110
16.0
110
16.0
110
14.5
110
18.0
110
20.5
110
20.0
110
18.0
100
0.0
120
25.0
110
Narrow Band Noise
ISO 389-4 1994
RETSPL
MaxHL
42.5
75
37.5
80
33.5
85
29.0
90
25.0
95
21.0
95
17.0
100
15.5
100
14.0
105
13.5
105
13.5
105
14.5
105
15.5
105
15.0
105
14.0
105
13.0
105
12.5
105
13.0
105
14.5
105
17.0
105
24.0
95
24.0
95
23.0
90
0.0
120

Hide quick links: