Interacoustics AD629 Instructions For Use Manual page 184

Diagnostic audiometer
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
ANSI HDA 200
Coupler: IEC60318-2 with type 1 adaptor
Tone Audiometer
Tone
ANSI S3.6-2010
Frequency
RETSPL MaxHL RETSPL 1/3 MaxHL
125
30.5
100
160¹
26.0
105
200¹
22.0
105
250
18.0
110
315¹
15.5
115
400¹
13.5
115
500
11.0
115
630¹
8.0
120
750
6.0
120
800¹
6.0
120
1000
5.5
120
1250¹
6.0
110
1500
5.5
115
1600¹
5.5
115
2000
4.5
115
2500¹
3.0
115
3000
2.5
115
3150¹
4.0
115
4000
9.5
115
5000¹
14.0
105
6000
17.0
105
6300¹
17.5
105
8000
17.5
105
9000
19.0
100
10000
22.0
100
11200
23.0
95
12500
27.5
90
14000
35.0
80
16000
56.0
60
18000²
72.0
40
20000²
86.0
25
WhiteNoise
¹ The RETSPL is copy from ISO389-8 2001.
² 18-20 kHz is estimated of Interacoustics.
ANSI EAR 3A
Coupler: ANSI S3.7-1995 (HA-2 with 5mm rigid Tube)
Tone Audiometer
Frequency
RETSPL
125
160
200
250
315
400
500
630
750
800
1000
1250
1500
1600
2000
2500
3000
3150
4000
5000
6000
6300
8000
WhiteNoise
Narrow Band Noise
ANSI S3.6-2010
34.5
75
30.0
80
26.0
80
22.0
85
19.5
90
17.5
95
15.0
95
13.0
95
11.0
100
11.0
100
11.5
100
12.0
95
11.5
100
11.5
100
10.5
100
9.0
100
8.5
100
10.0
100
14.5
100
19.0
95
22.0
90
22.5
90
22.5
90
24.0
85
27.0
85
28.0
80
32.5
75
40.0
70
61.0
50
77.0
25
91.0
10
0.0
115
Narrow Band
Tone
Noise
ANSI S3.6-2010
ANSI S3.6-2010
MaxHL
RETSPL
26.0
90
30.0
22.0
95
26.0
18.0
100
22.0
14.0
105
18.0
12.0
105
16.0
9.0
110
13.0
5.5
110
9.5
4.0
115
9.0
2.0
115
7.0
1.5
115
6.5
0.0
120
6.0
2.0
120
8.0
2.0
120
8.0
2.0
120
8.0
3.0
120
9.0
5.0
120
11.0
3.5
120
9.5
4.0
120
10.0
5.5
115
10.5
5.0
105
10.0
2.0
100
7.0
2.0
100
7.0
0.0
95
5.0
0.0
Coupler: IEC60318-1 with type 1 adaptor
Frequency
125
160
200
250
315
400
500
630
750
800
1000
1250
1500
1600
2000
2500
3000
3150
4000
5000
6000
6300
8000
8000
9000
10000
11200
12500
14000
16000
18000²
20000²
WhiteNoise
² 18-20 kHz is estimated of Interacoustics.
MaxHL
Frequency
90
95
100
105
105
105
110
110
110
110
110
110
110
110
110
110
110
110
110
105
100
100
95
110
WhiteNoise
IEC HDA 200
Tone Audiometer
Tone
Narrow Band Noise
ISO 389-8 2004
ISO 389-4 1994
RETSPL MaxHL
RETSPL 1/3
30.5
100
34.5
26.0
105
30.0
22.0
105
26.0
18.0
110
22.0
15.5
115
19.5
13.5
115
17.5
11.0
115
15.0
8.0
120
13.0
6.0
120
11.0
6.0
120
11.0
5.5
120
11.5
6.0
110
12.0
5.5
115
11.5
5.5
115
11.5
4.5
115
10.5
3.0
115
9.0
2.5
115
8.5
4.0
115
10.0
9.5
115
14.5
14.0
105
19.0
17.0
105
22.0
17.5
105
22.5
17.5
105
22.5
ISO 389-5 2006
ISO 389-5 2006 + 5 dB
17.5
105
22.5
19.0
100
24.0
22.0
100
27.0
23.0
95
28.0
27.5
90
32.5
35.0
80
40.0
56.0
60
61.0
72.0
40
77.0
86.0
25
91.0
IEC EAR 3A
Coupler: IEC 60318-5 2006 2ccm
Tone Audiometer
Tone
ISO 389-2 1994
RETSP
L
MaxHL
125
26.0
90
160
22.0
95
200
18.0
100
250
14.0
105
315
12.0
105
400
9.0
110
500
5.5
110
630
4.0
115
750
2.0
115
800
1.5
115
1000
0.0
120
1250
2.0
120
1500
2.0
120
1600
2.0
120
2000
3.0
120
2500
5.0
120
3000
3.5
120
3150
4.0
120
4000
5.5
115
5000
5.0
105
6000
2.0
100
6300
2.0
100
8000
0.0
95
MaxHL
75
80
80
85
90
95
95
95
100
100
100
95
100
100
100
100
100
100
100
95
95
95
90
90
85
85
80
75
70
50
25
10
0.0
115
Narrow Band Noise
ISO 389-4 1994
RETSPL MaxHL
30.0
90
26.0
95
22.0
100
18.0
105
16.0
105
13.0
105
9.5
110
9.0
110
7.0
110
6.5
110
6.0
110
8.0
110
8.0
110
8.0
110
9.0
110
11.0
110
9.5
110
10.0
110
10.5
110
10.0
105
7.0
100
7.0
100
5.0
95
0.0
110

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents