Safety Standards
EMC Standard
Audiometer Standards
Calibration
Air Conduction
Bone Conduction
Free Field
High Frequency
Effective masking
Transducers
Patient Response switch
Patient communication
Monitor
Special tests/test battery
Stimuli
Tone
Warble Tone
Wave file
AD629 Instruction for Use – English
Date: 2012-01-19
Technical Specifications
IEC 60601-1, ES60601-1, CAN/CSA-C22.2 No.60601-1
Class I, Applied parts type B, Continuous operation
IEC 60601-1-2:2001 + A1:2004
Tone Audiometer: IEC 60645 -1, ANSI S3.6 -2010, Type 2, HF IEC 60645-4.
Speech Audiometer: IEC 60645-2/ANSI S3.6 type B or B-E.
Auto threshold tests: ISO 8253-1
Calibration information and instructions is located in the AD629 Service manual
DD45:
TDH39:
HDA200:
HDA280
E.A.R Tone 3A/5A: ISO 389-2 1994, ANSI S3.6-2010
CIR 33
B71:
Placement:
ISO 389-7 2005, ANSI S3.6-2010
ISO 389-5 2004, ANSI S3.6-2010
ISO 389-4 1994, ANSI S3.6-2010
DD45
TDH39
HDA200
HDA280
B71 Bone
E.A.R Tone 3A/5A:
CIR 33
One hand held push button.
Talk Forward (TF) and Talk Back (TB).
Output through built-in speaker or through external earphone or speaker.
SISI. ABLB. Stenger. Stenger Speech. Langenbeck (tone in noise). Békésy Test.
2 channel speech, 2 channels Master Hearing Aid, Auto threshold.
Auto threshold tests:
Available time for patient to respond:
Increment of hearing level:
Auto threshold test (Békésy):
Mode of operation:
Rate of level change:
Smallest increment of level:
125-20000Hz separated in two ranges 125-8000Hz and 8000-20000Hz.
Resolution 1/2-1/24 octave.
1-10 Hz sine +/- 5% modulation
44100Hz sampling, 16 bits, 2 channels
PTB/DTU report 2009
ISO 389-1 1998, ANSI S3.6-2010
ISO 389-8 2006, ANSI S3.6-2010
PTB report 2004
ISO 389-2
ISO 389-3 1994, ANSI S3.6-2010
Mastoid
Headband Static Force 4.5N 0.5N
Headband Static Force 4.5N 0.5N
Headband Static Force 4.5N 0.5N
Headband Static Force 5N 0.5N
Headband Static Force 5.4N 0.5N
Same as tone presentation
5dB.
Békésy
2.5 dB/s ±20%
0.5 dB
Page 13/17