Calibration
DC current calibration
After the 1MΩ resistance point has been calibrated, the unit will prompt you to apply 10mA.
Follow these steps for DC current calibration:
1.
Figure 2-3
Connections for DC and AC amps calibration
Model 2700
Model 2700 Multimeter / Data Acquisition System
RATIO
CH AVG
CONT
OCOMP
MATH O U T P U T
Ω2
Ω4
SHIFT
DCV
ACV
DCI
ACI
HOLD
MONITOR
DELAY
LIMITS
ON/OFF
TYPE
LOCAL
EX TRIG
TRIG
STORE
RECALL
FILTER
REL
POWER
SAVE
SETUP
CONFIG
HALT
TEST
LSYNC
OPEN
CLOSE
STEP
SCAN
DIGITS RATE
2.
NOTE
Table 2-5
DC current calibration summary
Calibration step
10mA
100mA
1A
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2700 as
shown in
Figure
2-3.
Integra Series
SENSE
INPUT
Ω 4 WIRE
HI
350V
1000V
!
PEAK
PEAK
LO
500V
PERIOD SENSOR
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
CH-OFF
CARD
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
GPIB
RS-232
EXIT
ENTER
Calibrate each current step summarized in
Set the calibrator to the indicated DC current, and make sure the unit is in
OPERATE.
Make sure the Model 2700 display indicates the correct calibration current.
Press ENTER to complete each step.
Allow the Model 2700 to finish each step.
If you are performing DC-only calibration, proceed to "Setting calibration dates
and saving calibration."
Calibrator current
10.00000mA
100.0000mA
1.000000A
DC and AC Current Calibrator
Input
LO
Output HI
Amps
Output
Note: Be sure calibrator is set for
LO
normal current output.
Table
2-5. For each step:
Allowable range
9mA to 11mA
90mA to 110mA
0.9A to 1.1A