2-10
Calibration
DC current calibration
After the 1MΩ resistance point has been calibrated, the unit will prompt you to apply 10mA.
Follow these steps for DC current calibration:
1.
Figure 2-3
Connections for DC and AC amps calibration
Model 2700
Model 2700 Multimeter / Data Acquisition System
RATIO
CH AVG
CONT
OCOMP
MATH O U T P U T
Ω2
Ω4
SHIFT
DCV
ACV
DCI
ACI
HOLD
MONITOR
DELAY
LIMITS
ON/OFF
TYPE
LOCAL
EX TRIG
TRIG
STORE
RECALL
FILTER
REL
POWER
SAVE
SETUP
CONFIG
HALT
TEST
LSYNC
OPEN
CLOSE
STEP
SCAN
DIGITS RATE
2.
NOTE
Table 2-5
DC current calibration summary
Calibration step
10mA
100mA
1A
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2700 as
shown in
Figure
2-3.
Integra Series
SENSE
INPUT
Ω 4 WIRE
HI
350V
1000V
!
PEAK
PEAK
LO
500V
PERIOD SENSOR
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
CH-OFF
CARD
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
GPIB
RS-232
EXIT
ENTER
Calibrate each current step summarized in
•
Set the calibrator to the indicated DC current, and make sure the unit is in
OPERATE.
•
Make sure the Model 2700 display indicates the correct calibration current.
•
Press ENTER to complete each step.
•
Allow the Model 2700 to finish each step.
If you are performing DC-only calibration, proceed to "Setting calibration dates
and saving calibration."
Calibrator current
10.00000mA
100.0000mA
1.000000A
DC and AC Current Calibrator
Input
LO
Output HI
Amps
Output
Note: Be sure calibrator is set for
LO
normal current output.
Table
2-5. For each step:
Allowable range
9mA to 11mA
90mA to 110mA
0.9A to 1.1A