AC calibration commands
The :AC commands perform comprehensive (user) calibration of the ACV and ACI func-
tions.
Table B-3
Table B-3
AC calibration commands
Command
:CALibration
:PROTected
summarizes these calibration commands.
:AC
:STEP1
:STEP2
:STEP3
:STEP4
:STEP5
:STEP6
:STEP7
:STEP8
:STEP9
:STEP10
:STEP11
:STEP12
:STEP13
Description
10mV AC at 1kHz calibration step.
100mV AC at 1kHz calibration step.
100mV AC at 50kHz calibration step.
1V AC at 1kHz calibration step.
1V AC at 50kHz calibration step.
10V AC at 1kHz calibration step.
10V AC at 50kHz calibration step.
100V AC at 1kHz calibration step.
100V AC at 50kHz calibration step.
700V AC at 1kHz calibration step.
100mA AC at 1kHz calibration step.
1A AC at 1kHz calibration step.
2A AC at 1kHz calibration step.
Calibration Reference