Pulse Amplitude Index For The Ww1020 Oximeter; Pulse Amplitude Index For The Ww1000 Oximeter; Printer - BCI WW1020 Operation Manual

Smiths medical pulse oximeter operation manual
Table of Contents

Advertisement

Chapter 11: Specifications

Pulse Amplitude Index for the WW1020 Oximeter

Range:
Display:
Display Update Rate:
The oximeter will display dashes (Invalid data) for SpO
Low perfusion performance is verified by simulator testing.
As tested with Industry Standard Simulator.
6
The PI value maps to a 9-segment bar graph as shown below. The two lowest bars (1&2) are bi-
color (Yellow & Green), and bars 3 through 9 are Green.
PI From
0.00-
0.08-
Oximeter
0.07
Segments
1Y
Lit

Pulse Amplitude Index for the WW1000 Oximeter

Range:
Display:
Display Update Rate:
As tested with Industry Standard Simulator.
5
The PI value maps to a 9-segment bar graph as shown below. The two lowest bars (1&2) are bi-
color (Yellow & Green), and bars 3 through 9 are Green.
PI From
0.00
>0 - 0.2
Oximeter
Segments
1Y
Lit

Printer

Paper:
Modes:
Paper Speed:
Trend Graphs:
11-6
0.03% to 20.00%
(0.01% increments)
6
Pulse Amplitude Index is defined as PI = (100 * AC)/DC where AC is
the alternating current (pulsatile component of the signal) and DC is
direct current (non-pulsatile component of the signal). It is a relative
measure of pulse-signal strength over time at a pulse oximeter
monitoring site, and is non-pulsatile in nature.
9-segment green / yellow LED bargraph, logarithmic scale
10 Hz
0.16-
0.32-
0.15
0.31
0.63
2Y
3G
4G
0.00% to 20.00%
(0.2% increments)
5
Pulse Amplitude Index is defined as PI = (100 * AC)/DC where AC
is the alternating current (pulsatile component of the signal) and
DC is direct current (non-pulsatile component of the signal). It is
a relative measure of pulse-signal strength over time at a pulse
oximeter monitoring site, and is non-pulsatile in nature.
9-segment green / yellow LED bargraph, logarithmic scale
10 Hz
>0.2
>0.4
- 0.4
- 0.6
2Y
3G
4G
58mm (2.28 inches), Thermal
Real Time, Tabular Trend, Graphic Trend
12.5 mm/sec
Headers print with graphs
and PR when the PI value is below 0.03%.
2
0.64-
1.28-
2.56-
1.27
2.55
5.11
5G
6G
7G
>0.6
>1.2
>2.6
- 1.2
- 2.6
- 5.0
5G
6G
7G
Clinician's Operation Manual
5.12-
10.24-
10.23
20.47
8G
9G
>5.0
>10.2
- 10.2
8G
9G

Hide quick links: