Chapter 8: Maintenance

Cleaning and Disinfecting

WARNING! Do not autoclave, ethylene oxide sterilize, or immerse the monitor in liquid.
Clean the surfaces of the monitor with a soft cloth moistened in water or a mild soap solution. If
disinfecting is necessary, wipe the surfaces of the monitor with a 70% isopropyl alcohol solution.
If there is contamination with blood borne pathogens or other potentially infectious materials,
then the use of an approved disinfectant of appropriate spectrum for the suspected organisms is
appropriate. Then wipe the surfaces with a soft, water-moistened cloth.
WARNING! Do not allow water, isopropyl alcohol or any other liquid to enter any of the
openings on the monitor. Unplug the AC power cord from the monitor before
cleaning or disinfecting.
CAUTION! Where the equipment has accidentally gotten wet, it should be wiped dry
externally and allowed to dry thoroughly before use.
CAUTION! Before cleaning or disinfecting the printer, unplug the AC adapter, remove the
batteries and remove the paper.
CAUTION! Do not allow printer paper to become wet. If the printer paper gets wet, remove
the paper immediately. Do not use the printer until the paper is replaced.
CAUTION! Disinfectant chemicals may affect the outer case over prolonged use, however
disinfection must be performed.
NOTE! Use only a soft cotton cloth to clean the monitor's screen. Do not clean the screen
with tissues, paper towels, or any other paper-based wipe. Paper-based wipes can
scratch the screen.

Storage

Whenever possible, the monitor should be stored inside at room temperature in a dry
environment. If it is necessary to store the monitor for an extended period of time, the unit should
be packed in its original shipping container. Storing the monitor for a long period of time may
degrade the battery capacity. Batteries should be removed from the monitor before storing.
See Chapter 11 for storage specifications.
NOTE! The monitor may not meet performance specifications when stored or used outside
the temperature and humidity ranges listed in Chapter 11: Specifications.