Animal Weighing/Averaging

Display symbol: V
Purpose
Use this program to determine the weights of unstable
samples (e.g., live animals) or to determine weights under
unstable ambient conditions. With this program, the bal-
ance calculates the weight as the average of a defined
number of individual weighing operations (also referred to
as "subweighing operations").
Preparation
§ Select the Animal weighing application in the menu: see
"Configuration."
§ Set the following parameters:
Applic. Application program
AnimalW. Animal weighing
activty. Animal activity
Calm Stable readout
o Active Unstable
V.Active Very unstable readout
Start
manual
o Autom. Automatic
o = Factory setting
Changing the Number of Subweighing
Operations
Activate function:
Press the (MENU)* key
Select the desired number of
measurement (1 to 100):
In increments of 1:
Press the(MENU)* key briefly
In increments of 10:
Press and hold the (MENU)* key.
The selected number of measurements is stored in
battery-backed memory
Printout: Animal weighing
mDef
20
x-Net + 410.1 g
* On some models, the keypad overlay shows the German
word "MENUE" for MENU.
: Number of subweighing
operations
: Calculated average
33