Acculab ATL-224 Operating Instructions Manual page 8

Acculab atilon electronic analytical and precision balances
Table of Contents

Advertisement

8
Connecting Electronic Peripheral Devices
§ Make sure to unplug the balance from AC power before you connect or disconnect
a peripheral device (printer or computer) to or from the interface port.
Warmup Time
To ensure accurate results, the balance must warm up before operation
as follows:
– Models ATL-153 and ATL-822: 2 hours
– All other models: 30 minutes
Only after this time will the instrument have reached the required operating
temperature.
Using Verified Balances in Legal Metrology in the EU*:
$ Make sure to allow the equipment to warm up for at least 24 hours after initial
connection to AC power or after a relatively long power outage.
Operation Outside the Temperature Range
Operating the balance beyond the temperature range of +10...+30°C
!
(50°...86°F). Differences from the specifications listed in the chapter on
"Specifications" are possible.
* Including the Signatories of the Agreement on the European Economic Area