Acculab ATL-224 Operating Instructions Manual page 32

Acculab atilon electronic analytical and precision balances
Table of Contents

Advertisement

Example:
Calculating the weight per unit area of paper: An A4 sheet of paper is used in this example, with surface dimensions of 0.210 m +
2
0.297 m = 0.06237 m
Parameter settings:
Applic. Calc.-Method-Div. (menu code 2. 8. 1. 2)
Step
1. Tare the balance
2. Activate divisor input
3. Set the divisor (in this example: 0.06237):
Position the decimal point,
Enter numerals
4. Store the divisor and initialize the balance
The current divisor remains stored in
battery-backed memory until the setting
is changed
5. Weight per unit area:
Place an A4 sheet of paper
on the balance
6. If desired, print result
7. Toggle display between
weight and calculated value
8. Unload the balance
9. Repeat as needed, starting from Step 5
* On some models, the keypad overlay shows the German word "MENUE" for MENU.
32
. To determine the weight per unit area, the total weight is divided by the surface.
Key (or instruction)
(TARE)
(MENU)*
(ENTER),
5+ (MENU)*,
2+ (ENTER),
Repeatedly
(MENU)* or press and hold:
(ENTER), etc.
(ENTER)
(PRINT)
Repeatedly:
(MENU)*
Display/Data output
0.00 g
______0.
__.00000
__.06000
__.06237
0.0 o
+
Div
0.6237
79.7 o
+
*
Res
+
79.7 o
4.97 g *
+
79.7 o *
+
0.0 o
+
*

Hide quick links: