Example: Counting parts into a container
Parameter settings:
Applic.-Net-Tot (menu code 2. 5.)
Step
1. Place empty container on the balance.
2. Tare the balance
3. Add first component
4. Store component data
5. Add next component
6. Store component data
7. Weigh in further components
as desired
8. Fill to desired final value view
the current total weight value
9. Print total weight and clear
the component memory
* On some models, the keypad overlay shows the German word "MENUE" for MENU.
36
Key (or instruction)
(TARE)
(ENTER)
(ENTER)
Repeat steps
5 and 6
(MENU)*
(CLEAR)
Display/Data output
65.0 g
0.0 g
+ 120.5 g
+
0.0 g
NET
*
Comp 1+
120.5 g
+
70.5 g
NET
*
+
0.0 g
NET
*
Comp 2+
70.5 g
+ 191.0 g
*
+ 2117.5 g
T-Comp+
2117.5 g