Example : Counting parts of equal weight
Parameter: Applic. - Count. (menu code 2. 3.)
Step
1. Place empty container
on the balance
2. Tare the balance
3. Add reference sample quantity
to container (in this example:
20 pcs)
4. Changing the reference sample quantity:
5. Select reference sample quantity:
In increments of 1 (1, 2, 3, etc. to 100)
In increments of 10 (10, 20, etc. to 100)
6. Confirm selected reference sample quantity
and start application. The current reference
weight remains stored until a new reference
weight remains stored until a new reference
is set or the power supply is interrupted
7. Add desired number of pieces.
8. If desired, print quantity
9. Toggle display between
mean piece weight, weight,
quantity
10. Unload the balance
11. Repeat as needed, starting from step 7
12. Delete reference value
* On some models, the keypad overlay shows the German word "MENUE" for MENU.
28
Key (or instruction)
(TARE)
(MENU)*
Repeatedly: (MENU)*
Press briefly
(MENU)* press and hold
(ENTER)
(PRINT)
Repeatedly:
(MENU)*
(CLEAR)
Display/Data output
+
22.6
g
0.0
g
ref 10
pcs
ref 20
pcs
+
20
pcs
*
nRef
20 pcs
wRef
1.07 g
+
500
pcs
Qnt
+
500 pcs
1.07
g
*
G
+
g
535.0
*
+
500
pcs
*
21
pcs
*
0.0
g