Basic Weighing Function - Acculab ATL-224 Operating Instructions Manual

Acculab atilon electronic analytical and precision balances
Table of Contents

Advertisement

Basic Weighing Function

Features
– Taring the balance
– Printing weights
Preparation
§ Switch on the balance:
Press (ON/OFF)
§ Tare the balance, if necessary:
Press (TARE)
$ If necessary, change the configuration settings:
See the chapter entitled "Configuration"
$ If desired, load the factory settings:
See the chapter entitled "Configuration"
Additional Functions
$ Switching off the balance:
Press (ON/OFF)
$ Balance in standby mode:
the current time is displayed
12
Using Verified Balances as Legal Measuring
Instruments in the EU*:
The type-approval certificate for verification applies only to
non-automatic weighing instruments. For automatic
operation with or without auxiliary measuring devices, you
must comply with the regulations applicable to the place of
installation.
§ Before using the balance as a legal measuring instrument,
calibrate and adjust it at the place of use using the built-in
motorized calibration weight; for details, see
"Calibration/Adjustment" in this chapter.
$ The temperature range (°C) indicated on the verification
label must not be exceeded during operation.
* Including the Signatories of the Agreement on the
Eruopean Economic Area