Internal Calibration/Adjustment (Only on ATL...-I und ATL...-V models)
Set the following parameters:
SETUP - Waage - CAL.Just. - Cal.Int. (menu code 1.1. 9. 4)
The built-in motorized calibration weight is applied and removed automatically for internal calibration.
Step
1. Unload and tare the balance
2. Start calibration
The built-in weight is
applied
automaticallyg
3. Calibration/adjustment executed
4. The built-in weight is removed
Key (or instruction)
(TARE)
(CAL)
Display
0.0 g
g
CAL.Int.
CAL.RUN.
g
CAL.end
0.0 g
15