Acculab VICON Operating Instructions Manual

Acculab VICON Operating Instructions Manual

Electronic precision scales

Advertisement

Please contact QuickMedical for this manual in
languages other than English.
QuickMedical is an authorized distributor for
Acculab. See the full line of Acculab at:
http:www.quickmedical.com/acculab/
Operating Instructions | Betriebsanleitung | Mode d'emploi |
Istruzioni per l'uso | Manual de instrucciones
ACCULAB VICON
Electronic Precision Scales/Balances | Elektronische Präzisionswaagen | Balances électroniques d'analyse et de précision |
Bilance elettroniche di precisione | Balanzas electrónicas de precisión
98648-012-72

Advertisement

Table of Contents
loading

Summary of Contents for Acculab VICON

  • Page 1 Please contact QuickMedical for this manual in languages other than English. QuickMedical is an authorized distributor for Acculab. See the full line of Acculab at: http:www.quickmedical.com/acculab/ Operating Instructions | Betriebsanleitung | Mode d’emploi | Istruzioni per l’uso | Manual de instrucciones ACCULAB VICON Electronic Precision Scales/Balances | Elektronische Präzisionswaagen | Balances électroniques d’analyse et de précision |...
  • Page 2: Table Of Contents

    Installation – It is recommended to connect Acculab accessories and options, as these are optimally designed for use with your balance/scale. – Do not open the balance/scale housing as this may void the...
  • Page 3: Getting Started

    Getting Started Equipment Supplied Installation – Balance/scale with in-use cover Choose a location that is not subject to the following nega- – Weighing pan tive influences: – Plug-in AC adapter – Heat (heater or direct sunlight) – Drafts from open windows and doors Additional equipment with models VIC-612, VIC-412, –...
  • Page 4 Connecting the Installing the Battery Leveling the Balance/Scale Balance/Scale to AC Power (not for models VIC-303, VIC-123, (only for models VIC-303, VIC-4MG, VIC-2MG) VIC-123, VIC-4MG, VIC-2MG) § It is recommended to use only the included AC adapter for optimal performance and safety. §...
  • Page 5: Operation

    Operation Basic Weighing Function Description of the Keys Features – Zeroing the balance You can zero the balance/scale within the entire weighing range, up to the maximum capacity. Preparation § Switch on the balance/scale: press the (ON/OFF) key (ON/OFF) On/off key: switches the balance/scale on and §...
  • Page 6: Application Programs

    Application Programs Toggling between Weight Units With this application program you can toggle the display of a weight value back and forth between two weight units (see table below). Example: Toggle weight unit from pounds [lb] (application) to grams [g] (basic unit) Step Key (or instruction) Display...
  • Page 7: Counting

    Counting Purpose With the Counting program you can determine the number of parts or items. Example: Determine the number of uncounted parts; weigh in the selectable reference sample quantity (in this example: 20) Step Key (or instruction) Display 1. Select application program (ZERO) >...
  • Page 8: Weighing In Percent

    Weighing in Percent Purpose This application program allows you to obtain weight readouts in percent which are in proportion to a reference weight. Example: Determine an unknown percentage; store the weight on the balance/scale as the reference percentage (100%) Step Key (or instruction) Display 1.
  • Page 9: Hold" Display

    “Hold” Display Purpose “Holds” the displayed value; also, the display will be locked for 5 seconds after removing the sample from the pan. Example: Determine weight of oversized sample Step Key (or instruction) Display 1. Select application program (ZERO) > 2 sec 2.
  • Page 10: Totalizing

    Totalizing Purpose With this application program you can add up successive weight values exerding capacity of balance/scale. Example: Totalizing weight values Step Key (or instruction) Display 1. Select application program (ZERO) > 2 sec 2. Select Totalizing (F) repeatedly 3. Confirm setting (ENTER) Symbol “...
  • Page 11: Specific Gravity

    Specific Gravity Purpose Use this application program to determine the specific gravity of a sample. The result is displayed with one decimal place. Beaker and wire not included with balance/scale. Example: Determine the specific gravity of a solid. Step Key (or instruction) Display 1.
  • Page 12: Calibration/Span Adjustment

    Calibration/Span Adjustment Calibration is recommended after initial installation and each time the balance/scale is moved. Features The weight required for calibration/adjustment is displayed. Calibration/adjustment can be performed only when: – there is no load on the balance/scale, Standard calibration weights for selected models: to remove, –...
  • Page 13: Configuration (Setup Menu)

    Configuration (Setup Menu) To configure the balance/scale; i.e., adapt the balance/scale to individual requirements. Step Key (or instruction) Display 1. Switch off the balance/scale (ON/OFF) 2. Switch on the balance/scale and (ON/OFF) while all segments are displayed (ZERO) briefly Navigation in the Setup Menu Press briefly Press and hold (ENTER)
  • Page 14: Error Codes

    (ON/OFF) The weight readout is obviously wrong The balance/scale was not calibrated/ Calibrate/adjust the balance/scale adjusted before weighing Balance/scale not zeroed Zero the balance/scale If any other errors occur, contact your local Acculab customer service center.
  • Page 15: Overview

    Overview Specifications Model VIC-303 VIC-123 VIC-4MG VIC-2MG VIC-612 VIC-412 VIC-212 Weighing capacity Readability 0.001 0.001 0.005 0.005 0.01 0.01 0.01 Tare range (subtractive) ≤±g 0.004 Linearity 0.003 0.01 0.01 0.03 0.03 0.035 Operating temperature range 10°C to 30°C (273°K to 303°K; 50°F to 86°F) Stabilization time (average) Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels;...
  • Page 16: Accessories (Options)

    The operator shall be responsible for any modifications to you must comply with the provisions as specified in the Acculab equipment and must check and, if necessary, correct applicable regulations for installation in your country. these modifications.

This manual is also suitable for:

Vic-612Vic-711Vic-511Vic-412Vic-212Acculab vicon vic-612 ... Show all

Table of Contents