Replacing The Leg Actuator - Invacare ILIFTEE User Manual

Portable patient lift
Table of Contents

Advertisement

Available languages

Available languages

8.10 Replacing the Leg Actuator

1171892-A
1.
Remove the wire lock A that secures the leg actuator cable
to the mast.
2.
Disconnect the leg actuator cable B from the bottom of the
control box C.
Remove hitch pins D securing the pin E to the leg actuator
3.
F and mast bracket G.
4.
Remove the pin from the leg actuator and mast bracket.
5.
Perform the following to remove the leg actuator from the pivot
bracket H(Detail A):
WARNING!
Risk of Injury
Injury to personnel or damage to the patient lift
may occur.
– Ensure that there is sufficient room to turn the
patient lift on its side and that the floor area is
clear of debris.
a.
Turn the patient lift on its side.
Remove the hitch pin I securing the pin J to the pivot
b.
bracket.
c.
Remove the pin from the leg actuator and pivot bracket.
d.
Return the patient lift to the upright position.
6.
Slide the leg actuator out of the slot K in the base of the patient
lift (Detail B).
7.
Reverse STEPS 1- 5 to install the leg actuator.
Position leg actuator into the mast bracket and move the
legs to align the holes in the leg actuator with the holes in
the mast bracket.
Maintenance
43