Adjusting Cantilever Arm Pad Depth/Replacing Cantilever Arm Pad; Adjusting Depth; Replacing - Invacare Compass XE Owner's Operator And Maintenance Manual

Hide thumbs Also See for Compass XE:
Table of Contents

Advertisement

SECTION 6—ARMS
Adjusting Cantilever Arm Pad Depth/Replacing
Cantilever Arm Pad
NOTE: For this procedure, refer to FIGURE 6.12.

Adjusting Depth

1. Remove the mounting screw from the rear of the arm‐rest pad and self‐taping screw 
if full length arm pads.
2. Depending on the desired arm pad depth, reposition the cantilever slide tube as 
follows:
A. For Desk Length Arm Pads ‐ Secure to one of five positions.
B. For Full Length Arm Pads ‐ Secure into the first adjustment hole.
3. Reattach the arm pad/cantilever slide tube to the arm tube with existing hardware.
4. Repeat for the opposite side, if necessary.

Replacing

1. Remove the two mounting screws from the armrest pad.
2. Replace with new armrest pad.
3. Secure with existing two mounting screws.
DETAIL "A" - DESK LENGTH
ARM PADS
Mounting
Screw
(Rear of
Armrest
Cantilever
Slide Tube
DO NOT remove this
mounting screw unless
replacing armrest pad.
FIGURE 6.12 Adjusting Cantilever Arm Pad Depth/Replacing Cantilever Arm Pad
Compass™XE and Allegro®
DETAIL "B" - FULL LENGTH ARM PADS
Mounting Screw
(Rear of Armrest Pad)
Cantilever Arm
Self-Tapping Screw
Cantilever Slide Tube
DO NOT remove this mounting
screw unless replacing armrest pad.
NOTE: Only the circled adjustment holes can be used.