Replacing/Repositioning the Locking Mechanism in the
Cantilever Arm
NOTE: For this procedure, refer to FIGURE 6.8.
1. Move the cantilever arm up and out of the way.
2. Remove the actuator from the locking mechanism.
The locking mechanism is spring loaded. Place your free hand over the locking
mechanism to prevent the parts from springing out of the cantilever arm.
3. Slowly let the locking mechanism and spring slide out of the cantilever arm. 
NOTE: Inspect the spring and replace if necessary.
4. Slide the new locking mechanism and spring into the cantilever arm.
5. Position the angled portion of the locking mechanism in one of two ways:
A. Angled Portion Facing up ‐ Arm will flip up.
B. Angled Portion Facing down ‐ Arm will flip down.
6. Use Loctite 242
7. To lock the cantilever arm, push down until there is an audible click.
8. Pull up on the cantilever arm to make sure it is locked in place.
Angled Portion of
Locking Mechanism
FIGURE 6.8 Adjusting Cantilever Arm Pad Depth/Replacing Cantilever Arm Pad
Part No 1127114
CAUTION
 and securely tighten the actuator into the locking mechanism.
Locking
Mechanism
Actuator (Apply Loctite 242)
Spring
51
SECTION 6—ARMS
Cantilever Arm
Compass™XE and Allegro®